22 January 2016 Two-dimensional dynamic fluid bowtie attenuators
James R. Hermus, Timothy P. Szczykutowicz
Author Affiliations +
Abstract
Fluence field modulated (FFM) CT allows for improvements in image quality and dose reduction. To date, only one-dimensional modulators have been proposed, as the extension to two-dimensional (2-D) modulation is difficult with solid-metal attenuation-based fluence field modulated designs. This work proposes to use liquid and gas to attenuate the x-ray beam, as unlike solids, these materials can be arranged allowing for 2-D fluence modulation. The thickness of liquid and the pressure for a given path length of gas were determined that provided the same attenuation as 30 cm of soft tissue at 80, 100, 120, and 140 kV. Liquid iodine, zinc chloride, cerium chloride, erbium oxide, iron oxide, and gadolinium chloride were studied. Gaseous xenon, uranium hexafluoride, tungsten hexafluoride, and nickel tetracarbonyl were also studied. Additionally, we performed a proof-of-concept experiment using a 96 cell array in which the liquid thickness in each cell was adjusted manually. Liquid thickness varied as a function of kV and chemical composition, with erbium oxide allowing for the smallest thickness. For the gases, tungsten hexaflouride required the smallest pressure to compensate for 30 cm of soft tissue. The 96 cell iodine attenuator allowed for a reduction in both dynamic range to the detector and scatter-to-primary ratio. For both liquids and gases, when k-edges were located within the diagnostic energy range used for imaging, the mean beam energy exhibited the smallest change with compensation amount. The thickness of liquids and the gas pressure seem logistically implementable within the space constraints of C-arm-based cone beam CT (CBCT) and diagnostic CT systems. The gas pressures also seem logistically implementable within the space and tube loading constraints of CBCT and diagnostic CT systems.
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4302/2016/$25.00 © 2016 SPIE
James R. Hermus and Timothy P. Szczykutowicz "Two-dimensional dynamic fluid bowtie attenuators," Journal of Medical Imaging 3(1), 013502 (22 January 2016). https://doi.org/10.1117/1.JMI.3.1.013502
Published: 22 January 2016
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Liquids

Attenuators

Tissues

Signal detection

Modulation

Sensors

Signal attenuation

RELATED CONTENT

Fluid-filled dynamic bowtie filter: a feasibility study
Proceedings of SPIE (March 18 2015)
Fluid dynamic bowtie attenuators
Proceedings of SPIE (March 18 2015)
Optical Fiber Thermometer For Medical Use
Proceedings of SPIE (October 24 1984)
Electro-optic cosite interference mitigation
Proceedings of SPIE (February 23 2010)
Research of optical electric field probe
Proceedings of SPIE (October 15 2012)

Back to Top