Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters

[+] Author Affiliations
Kimberly Kontson

Center for Devices and Radiological Health, Food and Drug Administration, 10933 New Hampshire Avenue, Silver Spring, Maryland 20993, United States

University of Maryland, Department of Bioengineering, College Park, Maryland 20742, United States

Robert J. Jennings

Center for Devices and Radiological Health, Food and Drug Administration, 10933 New Hampshire Avenue, Silver Spring, Maryland 20993, United States

University of Maryland, Department of Bioengineering, College Park, Maryland 20742, United States

J. Med. Imag. 1(3), 033505 (Dec 18, 2014). doi:10.1117/1.JMI.1.3.033505
History: Received June 19, 2014; Accepted November 19, 2014
Text Size: A A A

Abstract.  Scatter contamination of projection images in cone-beam computed tomography (CT) degrades the image quality. The use of bowtie filters in dedicated breast CT can decrease this scatter contribution. Three bowtie filter designs that compensate for one or more aspects of the beam-modifying effects due to differences in path length in a projection were studied. These designs have been investigated in terms of their ability to reduce the scatter contamination in projection images acquired in a dedicated breast CT geometry. The scatter magnitude was measured as the scatter-to-primary ratio (SPR) using experimental and Monte Carlo techniques for various breast phantom diameters and tube voltages. The results show that a 55% reduction in the center SPR value could be obtained with the bowtie filter designs. On average, the bowtie filters reduced the center SPR by approximately 18% over all breast diameters. The distribution of the scatter was calculated at a range of different locations to produce scatter distribution maps for all three bowtie filter designs. With the inclusion of the bowtie filters, the scatter distribution was more uniform for all breast diameters. The results of this study will be useful in designing scatter correction methods and understanding the benefits of bowtie filters in dedicated breast CT.

Figures in this Article
© 2014 Society of Photo-Optical Instrumentation Engineers

Citation

Kimberly Kontson and Robert J. Jennings
"Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters", J. Med. Imag. 1(3), 033505 (Dec 18, 2014). ; http://dx.doi.org/10.1117/1.JMI.1.3.033505


Access This Article
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Book Chapters

Topic Collections

Advertisement
  • Don't have an account?
  • Subscribe to the SPIE Digital Library
  • Create a FREE account to sign up for Digital Library content alerts and gain access to institutional subscriptions remotely.
Access This Article
Sign in or Create a personal account to Buy this article ($20 for members, $25 for non-members).
Access This Proceeding
Sign in or Create a personal account to Buy this article ($15 for members, $18 for non-members).
Access This Chapter

Access to SPIE eBooks is limited to subscribing institutions and is not available as part of a personal subscription. Print or electronic versions of individual SPIE books may be purchased via SPIE.org.